Your position:Home > Products > Laboratory Equipments > Biological Safety Cabinet
Model |
BSC-1000B2 |
BSC-1300B2 |
BSC-1600B2 |
Overall size |
1200×800×2100 |
1500×800×2100 |
1800×800×2100 |
(W×D×H)mm |
(W×D×H)mm |
(W×D×H)mm |
|
Purification area size |
1000×700×580 |
1300×700×580 |
1600×700×580 |
(W×D×H)mm |
(W×D×H)mm |
(W×D×H)mm |
|
Purification efficiency |
The 100 stage(America federal standard 209E) |
||
Weight |
160kg |
180kg |
200kg |
Average air velocity |
0.38±0.025m/s Inside of the door |
0.27±0.025m/s In the side |
0.26±0.025m/s Middle |
Front inflow air |
0.48±0.025m/s |
||
Power |
0.8KW |
||
Average number of colonies |
≤0.5 /dish/hour) |
||
Noise |
≤65dB(A) |
||
Vibration/half peak amplitude |
≤3μm(X·Y·Z) |
||
Power supply |
220V 50HZ |
||
Filter size |
980×520×50 |
1280×520×50 |
1580×520×50 |
(W×D×H)mm |
(W×D×H)mm |
(W×D×H)mm |
|
Fluorescent lamp/UV lamp |
21W×②/21W×① |
28W×②/21W×① |
28W×②/21W×① |
Wind direction |
Top φ200mm |
||
Air tightness |
99.99≤E≤99.999 |
||
illumination |
≥700LX |