Your position:Home > Products > Clinical Analysis Equipments > Biochemistry Analyzer
FEATUES
Compact cabinet design-Space Efficiency
200 tests per hour-Time Efficiency
Automated testing & Low Maintenance-Operation Efficiency
150ul reagent consumption-Cost Efficiency
Getting the right answer at the first time-Laboratory Efficiency
Sampling probe
· External & internal mirror polish, external & internal probe washing(reducing carry-over)
· Dedicated sampling probe equipped with sensitive liquid sensor, timely feedback of reagent & sample residuals
· Collision protection
· Automatic probe depth adjusting
Multi-functional sample & reagent tray
· 80 reagent positions (40 for R1, 40 for R2)
· 40 sample positions, micro cup & test tube can be used
· Non-stop cooling system with peltier pad inside, 24 hours 2℃-14℃
· Bar code reader(optional)
、
Reaction tray
· 44 reaction cuvettes
· Reaction volume as low as 150ul
· Stable & accurate temperature(37±0.1 ℃) for reaction
High precision ceramic syringe
· Sample dispensing by frequency converting design
· Permanent ceramic piston
· Accurate dispensing as low as 0.1ul
Excellent mixer design
· Teflon coating mixer. No water dropping (reducing carry-over)
· Excellent mixing effect with standard mixing procedure
Washing station
· 4 needles wash-station
· Triple washing
· Effective washing to ensure accurate result and valid diagnostics
Powerful Software
· User friendly software interface simple, easy-to-operation
· Sample automatic dilution(decrease, increase, normal)
· Masking tests when reagent or standard is exhausted
· Calibrator auto-dilution
· Linear & nonlinear calibration
· Multi-points calibration up to 8 points
· Quality control apply to Westgard, L-T plot, Cumulative
· Real- time display of running status
· Real -time display of status of reagent tray, sample tray and reaction tray
· Real-time updating of reagent residuals
· Real-time marking of defective cuvette
· Reaction result automatic checking and warning
· Calibration result automatic checking and warning
· Alarm log: Automatic alarm and record in alarm list with explanations during running process
TECHNICAL SPECIFICATIONS
General Information
Machine type: Random access, open reagent & locked reagent (optional)
Test speed: Constant 200 tests/hour (mono reagent or double reagent)
Test principle: Colorimetric method, turbidimetry
Test method: 1 point end, 2 point end, Fixed-Time, Kinetics
Calibration type: Linear & nonlinear
Sample & Reagent Unit
Sample tray: 40 sample positions
Sample cup type: Micro cup & test tube
Reagent tray: 80 reagent positions(40 for R1 & 40 for R2)
Sample volume: 2-30ul, step by 0.1ul
Reagent volume: 20-300ul, step by 1ul
Sample & Reagent bar code reader(optional)
Reaction Unit
Reaction tray: 44 high quality plastic cuvettes
Reaction volume: 150ul-350ul
Reaction time: 10 minutes
Temperature: Peltier pad incubation system, reaction temperature (37℃±0.1 ℃)
Cuvette washing: 3*4 steps automatic washing
Mixing System
Teflon coating mixer
Optical System
Light source: Halogen lamp
Wavelength: 340nm; 405nm; 450nm; 505nm; 546nm; 578nm; 630nm; 700nm(4 more options)
Absorbance range: 0-4.0 Abs
Resolution: 0.0001 Abs
Operation System Requirements
Operation system: Windows XP & Windows 7
Interface: USB
Working Conditions
Power supply: AC 110/220V ± 10%, 50/60 Hz, 650W
Water consumption: 6L/hour(maximum)
Dimension: 650mm*700mm* 1200mm(L*W*H)