AJ-4102 100mA MEDICAL X-RAY Machine
Usage
Roentgenoscopy
Digestive track spot film photography
General photography
Filter-ray device photography
Features
1. Single bed and single bulb.
2. Design according to the electric mechanism of middle-size machine. The function is full and reasonable.
3. The high-tension system is of march-past structure. Equipped with full-wave rectification high-voltage generator and independent X-ray tube unit, this machine is with large output power and operation is reliable and stable.
4. Supply voltage V, roentgenoscopy, photography KV and step-less adjustable.
5. Equipped with the manostat for filament of X-ray tube and the space charge compensator as to ensure the stable and accurate output.
6. Equipped with the capacity limitation and protection circuit and KvmAS interlock limitation and protection.
7. Adopt the digital circuit timer, grade according to R10 priority coefficient and can control time exactly
8. Examining bed is of electric back rotation and is equipped with vibrating filter-ray device.
9. The spot film device of luminescent screen can hold the film jacket of various sizes and can make the photography of full-film and two sectional films separately
10. The upright post has the bottom track only, which is easy while installation.
Main specifications
ITEM |
CONTENT |
INDEX |
Upright post |
move lengthwise (along the bottom track) |
1500mm |
move transverse |
360mm |
move up and down (along upright post) |
720-1720mm(from focus to ground) |
Supply power |
voltage |
single phase 380V/220V |
frequency |
50Hz ± 0.5Hz |
capacity |
10KVA |
inner resistance |
380V:2 Ω ,220V:0.6 Ω |
Max. Dc output voltage 60KV 100KV ※ |
Max. Dc output current 100mA 125mA ※ |
Roentgenoscopy |
voltage |
50-90KV ※ 500-100KV※ |
current |
0.5-5mA |
Photography |
voltage |
50-90KV ※ 500-100KV※ |
current |
small focus 15mA, big focus 50, 60,100mA four grades together |
spot film |
100mA (fixed) |
time |
0.04-6.3s,24 grades together |
High-Voltage generator |
capacity |
100KVA |
X-Ray tube |
model |
XD4-2.9/100 |
focus |
small focus 1.8mm, big focus4.3mm |
Examining Bed |
rotary angle |
+90°- 0°- -5° |
specification of spot film(stomach and intestines photography) |
128X179mm(5”X7”) |
205X254mm(8”X10”) |
254X305mm(10”X12”) |
MAX.CAPACITY OF PHOTOGRAY
Focus of X-ray tube
|
photography mA |
Max. photography KV |
Max. allowable expose time |
small focus |
25mA |
90KV 100KV ※ |
1.0s |
big focus |
50 mA |
90KV 95KV ※ |
6.3s |
100 mA |
85KV 90KV ※ |
2.5s |
125 mA |
82KV 85KV ※ |
1.0s |
Note: As regard to the specification with ※, you can put forward your requirements when place order.
|