AJ-2204 Computerized high-frequency first-aid ventilator

This is a dedicated system device for respiration salvaging and revival, The main frame choose the principle of high-AJ-2204frequency ventilation, it is widely used for clinical salvaging and revival also is a dedicated device used for respiration salvaging on first-aid car.
Features
* .An opening system is designed to avoid leakage of airway, When ventilating, the intubation may without casing bag. Little tube may be used for ventilation when resistance of expiration is small, so that mucosa won't be hurt easily.
* With opening airway, suction and ventilation can be done at the same time, so good effectiveness of oxygen combination can be shown without inspiration and expiration synchronizing. 
* As ventilating, lower airway pressure is good to increase blood discharge and decrease encephal pressure.
* Suction device, it driving negative pressure is adjustable.
* Accumulator jar, it can be used continuously for  not less than four hours after being charging.
* The system disposes completely with laryngoscope, inlet intubation, catheter of tracheotomy, nasal prongs tube, air tube and so on.

Main Specifications    
* Output power: ≤65VA  Power supply: AC220V+22V 50HZ+1HZ
* Frequency:10~150 breaths/min
* I/E ratio:1:1, 1:1.5~1:3 and inverse ratio
* SIGH:6 breaths/min        * CPAP: Not less than 1.5KPa
* Suction negative pressure:-0.01~0MPa
* Driving Pressure:0~0.15MPa   Relative humidity : ≤80%    Temp.: 5-40%
* Pressure Regulator Limit:0.30MPa
* Constant volume ventilation: T=10min,I/E=1:1.5
Adults: f(100):f(20)=7:3  tidal volume=550ml/time   Driving pressure=0.06MPa
Children:  f(100):f(20)=7:3 tidal volume=350ml/time   Driving pressure=0.04MPa
* First-aid:   1) adult T=10min, F(100)=7min, F(20)=3min, Tidal=550ml/time, Cycle
2) Child T=10min, F(150)=7min, F(30)=3min, Tidal=350ml/time, Cycle
* Storage battery:4Ah,Automatic charge when 12V
* Most jetting volume: ≥1400ML/time
* Oxygen concentration: 35-90% adjustable
* Humidifying ratio: 0-15ml/hour
* Airway pressure display: 0-3Kpa
* Alarm: Alarm when battery pressure is lower than 11V;Alarm when the oxygen pressure is lower than 0.01MPa;Alarm with sound and light when no gas
* Ventilation Mode: HFJV, CPAP, HFJV+SIGH(Sigh), Constant volume ventilation.